The four-and-a-half digit readout provides a range of 1-9,999 Ω. In addition, its rechargeable battery allows up to 10 hours of mobile use. Now, the EVOM2™ is always on when you need it. The isolated power source of the EVOM2™ was specifically designed to avoid adverse effects on tissue and the formation of electrode metal deposits, even when it is plugged into a standard wall outlet. Isolated battery power for 10 hours of use When combined with WPI’s Endohm chamber, the EVOM2™ can also be used to perform more accurate quantitative measurements or lower resistance measurements like transendothelial electrical resistance measurements. The unique electronic circuit of the EVOM2™ and the included STX2 electrode detect the confluence of the cellular monolayer. The EVOM2™ not only qualitatively measures cell monolayer health, but also quantitatively measures cellular confluence. EVOM2™ is the next generation, redesigned for ease of use. The EVOM was the first instrument designed specifically to perform routine Trans Epithelial Electrical Resistance (TEER) measurement in tissue culture research. Making TEER Measurements to Determine Cellular Confluence TEER and trans-epithelial voltage measurements in 2D cell cultures.A variety of accessory electrodes are available for measuring TEER in 6- and 96-well fixed (HTS) and removable well culture systems (See STX100 series and Endohm electrodes).You can verify performance and calibrate the meter for TEER function using provided test resistor.Free standing with tilt bail, making viewing results easy.STX2 manual electrodes and test electrode included with every meter.Electrically isolated meter that plugs into a standard outlet for continual readout without push buttons.